PATIENT MONITOR
iM8 MONITOR
FEATURES:
SEAMLESS DESIGN
-
HIGH RESOLUTION 12" TFT-LCD SCREEN
-
LARGE FONT & CUSTOMIZABLE SCREEN DISPLAY
-
ROLLING STAND & WALL MOUNT OPTION

PULSE OXIMETER
m3A VITAL SIGN
FEATURES:
-
Dimensions: 208mm (L) x 241mm (H) x 189mm (D)
-
Weight: 3kg
-
Screen: 3.5 "color TFT-LCD, 640x480
-
Power Supply: 100-240 VAC, 50 / 60HZ
-
Battery: Lithium Li
-
Voltage: 14.8V DC
-
Capacity: 2200mAh / 4400 mAh
-
Maximum operating time: 17 hours (4400 mAh)
SpO2:
-
Measurement and alarm range: 0 ~ 100%
-
Resolution: 1%
-
Accuracy: ± 2% (70 ~ 100% / Adult / Pediatric) and ± 3% (70 ~ 100% / Newborn)
-
PR measurement and alarm range: 25 ~ 300 bpm
-
Resolution: 1 bpm
-
Accuracy: 3 bpm
-
Data update period: 1s
NIBP
-
Method: Automatic Pulse Meter
-
Measurement of systolic/diastolic/mean pressure
-
Measuring interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480 Min
-
Operating modes: Automatic operation, Manual operation, Continuous operation
Adult pressure range
-
Systolic: 40 ~ 270mmHg
-
Diastolic: 10 ~ 215mmHg
-
MAP: 20 ~ 235mmHg
Children pressure range
-
Systolic: 40 ~ 200mmHg
-
Diastolic: 10 ~ 150mmHg
-
MAP: 20 ~ 165mmHg
Newborn pressure range
-
Systolic: 135mmHg
-
Diastolic: 10 ~ 100mmHg
-
MAP: 20 ~ 110mmHg
-
Maximum mean error: 5mmHg
-
Maximum standard deviation: 8mmHg
-
Measuring period 40 ~ 240 bpm
-
Resolution: 1 bpm
-
Accuracy: 3bpm or 3.5%

FETAL MONITOR
F9 FETAL MONITOR
FEATURES:
-
Basic Parameters: FHR, TOCO, Event Mark, AFM.
-
Internal Parameters: IUP, DECG (optional).
-
Maternal Parameters: MECG, NIBP, MSpO2.
-
Display: 12.1" (30.7 cm) color TFT-LCD screen (touchscreen optional).
-
Display Resolution: 800 x 600.
-
Dimensions: 13.6" x 13" x 5" (347 mm x 330 mm x 126 mm).
-
Weight: 13.2 lb (6 kg).
-
Battery: 14.8 V rechargeable Lithium-Ion battery.
-
Battery Continuous Working Time: 4 hours.
-
Ultrasound:
-
Technique: Ultrasound Pulse Doppler with correlation.
-
Pulse Repetition Rate: 2 KHz.
-
Ultrasound Frequency: 1.0 MHz + 10%.
-
FHR Range: 50 bpm - 240 bpm.
-
Resolution: 1 bpm.
-
Accuracy: ±2 bpm.
-
-
DECG:
-
Technique: Peak - Peak detection technique.
-
FHR Range: 30 bpm - 240 bpm (American Standard), 50 bpm - 210 bpm (International Standard).
-
Resolution: 1 bpm.
-
Accuracy: ±1 bpm.
-
Input Impedance: > 10 M (Differential).
-
Input Impedance: > 20 M (Common Mode).
-
-
TOCO:
-
TOCO Range: 0 - 100 Relative (%).
-
Resolution: 1%.
-
Zero Mode: Auto/Manual.
-
Non-Linear Error: ±1 mmHg.
-
-
IUP:
-
Pressure Range: 0 - 100 mmHg.
-
Resolution: 1%.
-
Zero Mode: Auto/Manual.
-
Non-Linear Error: ±mmHg.
-
-
Auto Fetal Movement (AFM):
-
Technique: Pulsed Doppler Ultrasound.
-
Range: 0 - 100 (%).
-
Resolution: 1%.
-
-
Marking: Manual fetal movement mark.
-
SpO2:
-
Measuring Range: 0 - 100 Relative (%).
-
Resolution: 1%.
-
Measuring Accuracy: ±2 bpm.
-
-
NIBP:
-
Systolic Pressure: 50 mmHg - 240 mmHg.
-
Mean Pressure: 25 mmHg - 200 mmHg.
-
Diastolic Pressure: 15 mmHg - 180 mmHg.
-
Resolution: 1 mmHg.
-
-
Temperature:
-
Measuring Range: 82.4°F - 111.2°F (28°C - 44°C).
-
Accuracy: ±0.2°C.
-